Development and Validation of a Reversed-Phase High-Performance Liquid Chromatographic Method for the Simultaneous Determination of Amiloride Hydrochloride, Atenolol, Hydrochlorothiazide, and Chlorthalidone in Their Combined Mixtures
Faculty
Pharmacy
Year:
2009
Type of Publication:
Article
Pages:
404-409
Authors:
Elgawish, Mohamed S, Mostafa, Samia M, Elshanawane, Abdalla A
Journal:
JOURNAL OF AOAC INTERNATIONAL AOAC INT
Volume:
92
Research Area:
Chemistry; Food Science \& Technology
ISSN
ISI:000265309100007
Keywords :
Development , Validation , , Reversed-Phase High-Performance Liquid Chromatographic
Abstract:
A high-performance liquid chromatographic method was developed for the simultaneous determination of 2 ternary mixtures containing amiloride hydrochloride, atenolol, hydrochlorothiazide, and chlorthalidone used in hypertension therapy. The use of cyanopropyl column results in satisfactory separation of both mixtures. The mobile phase consisted of 10 mM KH(2)PO(4) buffer (pH 4.5) and methanol in a ratio of (75 + 25\% v/v), at a flow rate of 1 mL/min. UV detector was operated at 275 nm. Calibration graphs were linear in the concentration ranges of 2-10, 20-200, 10-100, and 5-50 mu g/mL for amiloride hydrochloride, atenolol, hydrochlorothiazide, and chlorthalidone, respectively. Intraday and interday precision values (relative standard deviation) were <1.13 for mixture I (amiloride hydrochloride, atenolol, chlorthalidone), and <0.93 for mixture II (amiloride hydrochloride, atenolol, hydrochlorothiazide). The method was successfully applied for the determination of the 2 combinations in laboratory-prepared mixtures and commercial pharmaceutical formulation with high accuracy and precision. Statistical comparison of the results with those of the published methods showed excellent agreement and indicates no significant difference between them.
Online
PDF
جامعة المنصورة
جامعة الاسكندرية
جامعة القاهرة
جامعة سوهاج
جامعة الفيوم
جامعة بنها
جامعة دمياط
جامعة بورسعيد
جامعة حلوان
جامعة السويس
شراقوة
جامعة المنيا
جامعة دمنهور
جامعة المنوفية
جامعة أسوان
جامعة جنوب الوادى
جامعة قناة السويس
جامعة عين شمس
جامعة أسيوط
جامعة كفر الشيخ
جامعة السادات
جامعة طنطا
جامعة بنى سويف