Anti-Infection Dip Suggestions for the Coloplast Titan Inflatable Penile Prosthesis in the Era of the Infection Retardant Coated Implant
Faculty
Not Specified
Year:
2011
Type of Publication:
Article
Pages:
2647-2654
Authors:
Wilson, Steven K, Salem, Emad A, Costerton, William
DOI:
10.1111/j.1743-6109.2011.02363.x
Journal:
JOURNAL OF SEXUAL MEDICINE WILEY-BLACKWELL
Volume:
8
Research Area:
Urology \& Nephrology
ISSN
ISI:000294561800026
Keywords :
Inflatable Penile Prosthesis, Implant Infection, Infection Retardant Coatings, Biofilm, Antibiotics
Abstract:
Introduction. Infection is the worst complication seen with inflatable penile prosthesis (IPP). Both the American Medical Systems (AMS) and Coloplast IPP have infection retardant coatings. AMS is coated at the factory with rifampicin and minocycline (InhibiZone). The Coloplast IPP has a hydrophilic coating covalently bonded to its components that will absorb any aqueous solution before implantation and provides increased surface lubricity to decrease bacterial adherence. Aim. We tested several antibiotic dips comparing zones of inhibition (ZOI) against five commonly infecting bacteria with coated Coloplast implants. Results were compared with those ZOI created with strips of an AMS IPP precoated with InhibiZone. Methods. Pieces of sterile Coloplast Titan IPP were dipped in (i) trimethoprim/polymixin B ophthalmic solution; (ii) trimethoprim/sulfamethoxazole infusion solution; (iii) bacitracin; (iv) rifampicin/minocycline; and (v) rifampin/trimehtoprim/sulfamethoxazole. ZOI for the Titan strips and for AMS InhibiZone coated strips were tested against Staphylococcus epidermidis, Staphylococcus lugdunensis, Staphylococcus aureus, Pseudomonas, and Enterococcus. Main Outcome Measure. ZOIs of the Coloplast Titan for each of the medicated solutions were compared with ZOI created by undipped strips of a sterile InhibiZone coated IPP placed on plates of the identical bacteria. Results. All dips except bacitracin showed ZOI >= InhibiZone (P >= 0.005) for most organisms. Because of broad-spectrum effectiveness, ease of handling, and cost, infusion vial of trimehtoprim/sulfamethoxazole seemed optimal at this time. If trimehtoprim/sulfamethoxazole is unavailable; the ZOI with Polytrim ophthalmic solution zones were almost as good. Conclusions. The Coloplast strips when dipped in several solutions showed equal or significantly larger ZOI against commonly infecting organisms than the InhibiZone coated strips. At the present time using off the shelf trimethoprim sulfamethoxazole infusion solution seems optimum. The flexibility of choosing the drug eluting from the Coloplast device seems promising in the changing bacterial environment. Wilson SK, Salem EA, and Costerton W. Anti-infection dip suggestions for the coloplast titan inflatable penile prosthesis in the era of the infection retardant coated implant. J Sex Med 2011; 8: 2647-2654.
Online
PDF
جامعة المنصورة
جامعة الاسكندرية
جامعة القاهرة
جامعة سوهاج
جامعة الفيوم
جامعة بنها
جامعة دمياط
جامعة بورسعيد
جامعة حلوان
جامعة السويس
شراقوة
جامعة المنيا
جامعة دمنهور
جامعة المنوفية
جامعة أسوان
جامعة جنوب الوادى
جامعة قناة السويس
جامعة عين شمس
جامعة أسيوط
جامعة كفر الشيخ
جامعة السادات
جامعة طنطا
جامعة بنى سويف