POTENTIOMETRIC DETERMINATION OF EPHEDRINE HYDROCHLORIDE USING PLASTIC MEMBRANE ION-SELECTIVE ELECTRODE

Faculty Science Year: 1993
Type of Publication: Article Pages: 371-376
Authors:
Journal: JOURNAL OF CHEMICAL TECHNOLOGY AND BIOTECHNOLOGY JOHN WILEY \& SONS LTD Volume: 58
Research Area: Biotechnology \& Applied Microbiology; Chemistry; Engineering ISSN ISI:A1993ML54400009
Keywords : EPHEDRINE, PHENYLEPHRINE TETRAPHENYLBORATE, PVC MEMBRANE ELECTRODE, POTENTIOMETRIC TITRATIONS    
Abstract:
A plastic membrane electrode based on phenylephrine-tetraphenylborate ion association shows a Nernstian response to both ephedrinium (Ep+) and silver (Ag+) cations over the concentration ranges 2 x 10(-5)-10(-1) m and 10(-5)-10(-1) m for Ep+ and Ag+, respectively. The performance characteristics of the electrode were investigated. The electrode proved to be an excellent potentiometric sensor for the determination of Ep-hydrochloride in its pure solutions and pharmaceutical preparations using the extrapolation method and potentiometric titration using a standard solution of AgNO3.
   
  Online    
PDF  
       